Agilent Injection Ports
The Microseal fits directly onto the septum cup of the Agilent Split/Splitless injection port. It requires a Microseal nut.
Recommendations for Use:
Merlin Microseal is compatible with all Agilent autosamplers & Split/Splitless ports (including 5890, 6890, 7890, 8890, and Intuvo series).
Tighten the Microseal nut finger tight only. Do not overtighten.
Injection port temperature range: 50- 400 C.
Requires a blunt, cone tip, 23 gauge syringe or 23 gauge SPME probes.
Microseal #410 is recommended for most applications (3 – 100 psi).
Use Microseal #310 for lower injection port pressures (1 – 45 psi).
Use Microseal # 21-01 Microseal for SPME.